One way of contributing to a demonstration that a medical device is acceptably safe is to show that the device satisfies a set of requirements known to mitigate hazards. This paper describes experience using formal techniques to model an IV infusion device and to prove that the modelled device captures a set of requirements. The requirements chosen for the study are based on a draft proposal developed by the US Food and Drug Administration (FDA). A major contributor to device related errors are (user) interaction errors. For this reason the chosen models and requirements focus on user interface related issues.
@InProceedings{HarrisonMCC:2014,
author = {M.D. Harrison and P. Masci and J.C. Campos and P. Curzon},
title = {Demonstrating that medical devices satisfy user related safety requirements},
booktitle = {Software Engineering in Healthcare (FHIES/SEHC 2014)},
series = {Lecture Notes in Computer Science},
volume = {9062},
pages = {113-128},
publisher = {Springer},
year = {2017},
doi = {10.1007/978-3-319-63194-3_8},
abstract = {One way of contributing to a demonstration that a medical device is acceptably safe is to show that the device satisfies a set of requirements known to mitigate hazards. This paper describes experience using formal techniques to model an IV infusion device and to prove that the modelled device captures a set of requirements. The requirements chosen for the study are based on a draft proposal developed by the US Food and Drug Administration (FDA). A major contributor to device related errors are (user) interaction errors. For this reason the chosen models and requirements focus on user interface related issues.},
paperurl = {http://www.di.uminho.pt/~jfc/publications/HarrisonCC17-FHIES2014-postprint.pdf}
}Generated by mkBiblio 2.6.28